„BioAir S.p.A” Safemate EZ series mikrobiologinė saugos spinta

  • Technical Specifications:
    • Fully complies with EN12469 requirements.
    • Latest generation microprocessor control system.
    • Large digital display.
    • Air and aerosol-tight push-button door, electrically operated with the touch of a finger.
    • Alarm for low airflow and incorrect front window position.
    • Sloped front and back wall for easy access.
    • Front access for filter maintenance and service.
    • C-shaped support with adjustable height.
    • Update rolls are usually installed.
    • HPV is compatible with sterilization.
  • Safemate EZ Cabinets are available in two sizes (1.2 m and 1.8 m).
    These latest generation Microbiological Safety Cabinets Class II Type A2 are designed to meet the strictest safety standards (EN12469-2000).
    Internal design, airflow aerodynamics and monitoring, integrated safety devices and a highly precise manufacturing process ensure the highest results, meeting EN12469 standards.
  • High biosafety combined with competitive prices provides the consumer with a new standard of affordable budgets that only experienced European design and precise quality manufacturing processes can offer.
    The EZ Series sets a new standard for entry-level cabinets, combining a very attractive price with a full range of convenience features that were previously only available in more expensive cabinets.
  • Main Specifications:
    • Fully complies with EN12469 requirements.
    • Microprocessor-controlled fan with a volumetric sensor for the extracted air flow.
    • The latest generation microprocessor control system offers:
      • Big screen.
      • Automatic control of set air flow rates.
      • A sliding door window with smart control.
      • Continuous monitoring of the HEPA filter's service life.
      • Alarm. Multi-function alarm with backup functions.
      • Continuous display of working conditions.
      • High air flow stability in both transient disturbances and progressive filter clogging.
      • Continuous monitoring of the front barrier airflow to guarantee operator safety.
      • Low barrier signal.
    • Removing the electric fan from the rear.
    • Automatic reset to the original state if power fails.
    • C-shaped support with adjustable height for easy one-person installation.
  • Mechanical and Functional Specifications:
    • 5° Sloped front design increases working comfort. Sloped rear wall of the working chamber for best downward flow distribution.
    • Utility entries from above into the closet.
    • Stalinės plieninės vidinės paviršiai su SB apdaila (įskaitant nutekėjimo padėklą). Stalas arba tinklelio struktūros darbinė paviršius (padalintas sektorais) ir „V” formos priekinė grotelė su nuobodžiu kliūčių priekinėje grotelėje.
    • Electrically operated sliding laminated safety glass windows (maximum opening 120°).
    • Convenient 200 mm front opening.
    • Easy to install upgrade options through the side edges.
    • External exhaust HEPA filter for easy visual integrity inspection.
    • H14 klasės aukšto našumo dalelių filtras su 99,999% efektyvumu, atitinkančiu 0,3 mikrono dalelių (daugiausia skverbiasi dalelių skersmuo) (efektyvumas >= 99,995% 0,1-0,2 mikrono dalelėms MPPS pagal EN1822-1).
    • The intake, exhaust, and main filters have a micromesh membrane located at the end of the downstream equalizer plenum, acting as expansion chambers for the air velocity equalizer and a clear indicator of filter damage.
    • Filter replacement and maintenance from the front.
    • Exhaust transitions are easy to install.
    • Key operation. The key can be removed when the unit is in SAFE mode to prevent unwanted operation. In the event of a power failure, the cabinet is returned to its original condition.
    • Automatic self-adjustment cycle when the cabinet is turned on.
    • Visual indication of SAFE status. Warning (visual and audible signals) before reaching the actual alarm status.
    • Soft touch control keys with keys for standard service tools. Interlinked UV and fluorescent lights.
    • The microprocessor is equipped with an analog watchdog.
  • Default Utilities:
    • UV lamp on the back wall. Control panel with delayed start and cycle duration (max. 1 hour).
    • Two electrical plugs (Schucko/Europlug standard). Other plug standards available as options.
    • Lamps without fluorescent lamps.
  • Additional Utilities:
    • Compressed air/vacuum tube. Installation on the right wall.
    • Gas plug with electromagnetic safety valve. Installation on the right wall.
    • HPV adapter kit (reception plug on the left wall and suction plug connection for transition adapter).
    • Additional plugs and plug standards (maximum power up to 3 amps).
    • Inactive and active transition adapters or for peak ventilation.

Mikrobiologinės Saugos Spintos – BioAir laboratory furniture group

More information: on the manufacturer's website or send us an inquiry!

Manufacturer

BioAir SpA

Device power

1200W, 1750W

Traction speed (EN 12469)

from 0.35 to 0.40 ±10% m/s

Traction performance

450 m3/h ±10%, 600 m3/h ±10%

Power supply

230V @ 50Hz

Operating ambient temperature

from +10ºC to +35ºC

Product dimensions (length × depth × height)

1.360-1.990 × 0.780 × 1.450 m

Product weight (net)

130-185kg

en_GBEN