„BioAir SpA Safemate EZ series microbiological safety cabinet

  • Technical Specifications:
    • Fully complies with EN12469 requirements.
    • Latest generation microprocessor control system.
    • Large digital display.
    • Air and aerosol-tight push-button door, electrically operated with the touch of a finger.
    • Alarm for low airflow and incorrect front window position.
    • Sloped front and back wall for easy access.
    • Front access for filter maintenance and service.
    • C-shaped support with adjustable height.
    • Update rolls are usually installed.
    • HPV is compatible with sterilization.
  • Safemate EZ Cabinets are available in two sizes (1.2 m and 1.8 m).
    These latest generation Microbiological Safety Cabinets Class II Type A2 are designed to meet the strictest safety standards (EN12469-2000).
    Internal design, airflow aerodynamics and monitoring, integrated safety devices and a highly precise manufacturing process ensure the highest results, meeting EN12469 standards.
  • High biosafety combined with competitive prices provides the consumer with a new standard of affordable budgets that only experienced European design and precise quality manufacturing processes can offer.
    The EZ Series sets a new standard for entry-level cabinets, combining a very attractive price with a full range of convenience features that were previously only available in more expensive cabinets.
  • Main Specifications:
    • Fully complies with EN12469 requirements.
    • Microprocessor-controlled fan with a volumetric sensor for the extracted air flow.
    • The latest generation microprocessor control system offers:
      • Big screen.
      • Automatic control of set air flow rates.
      • A sliding door window with smart control.
      • Continuous monitoring of the HEPA filter's service life.
      • Alarm. Multi-function alarm with backup functions.
      • Continuous display of working conditions.
      • High air flow stability in both transient disturbances and progressive filter clogging.
      • Continuous monitoring of the front barrier airflow to guarantee operator safety.
      • Low barrier signal.
    • Removing the electric fan from the rear.
    • Automatic reset to the original state if power fails.
    • C-shaped support with adjustable height for easy one-person installation.
  • Mechanical and Functional Specifications:
    • 5° Sloped front design increases working comfort. Sloped rear wall of the working chamber for best downward flow distribution.
    • Utility entries from above into the closet.
    • Tabletop steel internal surfaces with SB finish (including drain pan). Table or grid structure work surface (divided into sectors) and "V" shaped front grille with boring of obstacles in the front grille.
    • Electrically operated sliding laminated safety glass windows (maximum opening 120°).
    • Convenient 200 mm front opening.
    • Easy to install upgrade options through the side edges.
    • External exhaust HEPA filter for easy visual integrity inspection.
    • H14 class high-performance particulate filter with 99.999% efficiency corresponding to 0.3 micron particles (most penetrating particle diameter) (efficiency >= 99.995% for 0.1-0.2 micron particles MPPS according to EN1822-1).
    • The intake, exhaust, and main filters have a micromesh membrane located at the end of the downstream equalizer plenum, acting as expansion chambers for the air velocity equalizer and a clear indicator of filter damage.
    • Filter replacement and maintenance from the front.
    • Exhaust transitions are easy to install.
    • Key operation. The key can be removed when the unit is in SAFE mode to prevent unwanted operation. In the event of a power failure, the cabinet is returned to its original condition.
    • Automatic self-adjustment cycle when the cabinet is turned on.
    • Visual indication of SAFE status. Warning (visual and audible signals) before reaching the actual alarm status.
    • Soft touch control keys with keys for standard service tools. Interlinked UV and fluorescent lights.
    • The microprocessor is equipped with an analog watchdog.
  • Default Utilities:
    • UV lamp on the back wall. Control panel with delayed start and cycle duration (max. 1 hour).
    • Two electrical plugs (Schucko/Europlug standard). Other plug standards available as options.
    • Lamps without fluorescent lamps.
  • Additional Utilities:
    • Compressed air/vacuum tube. Installation on the right wall.
    • Gas plug with electromagnetic safety valve. Installation on the right wall.
    • HPV adapter kit (reception plug on the left wall and suction plug connection for transition adapter).
    • Additional plugs and plug standards (maximum power up to 3 amps).
    • Inactive and active transition adapters or for peak ventilation.

Microbiological Safety Cabinets – BioAir laboratory furniture group

More information: on the manufacturer's website or send us an inquiry!

Manufacturer

BioAir SpA

Device power

1200W, 1750W

Traction speed (EN 12469)

from 0.35 to 0.40 ±10% m/s

Traction performance

450 m3/h ±10%, 600 m3/h ±10%

Power supply

230V @ 50Hz

Operating ambient temperature

from +10ºC to +35ºC

Product dimensions (length × depth × height)

1.360-1.990 × 0.780 × 1.450 m

Product weight (net)

130-185kg

en_GBEN